NI2000PAT On-line Near-Infrared Analyzer
Multi-channel options: up to 8-channel simultaneous detection for higher cost-effectiveness
High sensitivity: cooled detector with detection limit below 0.1%
High speed: millisecond-level rapid response
High flexibility: compatible with various classic diffuse-reflectance probes, transmission-reflection probes or transmission probes of different optical path lengths, as well as flow cells with multiple specification interfaces
Powerful functions: real-time monitoring of multiple components and their content variations
Versatile testing accessories: immersion probe for on-line monitoring; optional diffuse-reflectance reference plate, fine-tunable probe mounting bracket, closed detection chamber, etc. Suitable for testing solids, powders, liquids and other sample forms.
1. Agriculture and Food Industry
2. Chemical & Pharmaceutical Industry, Fine Biocatalysis
3. Textile Industry
4. Environmental Monitoring
5. Medical Care and Biomedicine
6. Petrochemical Industry
7. Plastics and Polymers
8. Archaeology and Art Conservation
Glucose aqueous solution quantitative modeling; error < 0.1%; repeatability < 0.1%
Online Monitoring
Spectral Data Acquisition
Trend Display
Model Analysis
|
PerformanceItem |
Parameter |
| Application Mode-Testing Accessory | Immersion Probe |
| Application Mode-Channel | Standard: single-channel; 8-channel optional |
| Optical Parameter-Spectral Range | 900 ~ 1700 nm or 900 ~ 2500 nm (optional) |
| Optical Parameter-Spectral Resolution | 3.1 nm @ 900 ~ 1700 nm; 6.3 nm @ 900 ~ 2500 nm |
| Optical Parameter-Detector | Cooled linear-array InGaAs |
| Optical Parameter-Light Source | 20 W@2500 hours; 5 W@10000 hours; quick-detachable and replaceable |
| Immersion Probe-Mode | Standard: transmission-reflection; other modes available. Compatible with diverse reaction kettles and reaction conditions |
| Immersion Probe-Temperature Rating | < 150 °C; customized for higher conditions |
| Immersion Probe-Pressure Rating | < 1 MPa; customized for higher conditions |
| Immersion Probe-Probe Length | Standard: 300 mm; customizable length |
| Immersion Probe-Probe Diameter | 12 mm |
| Immersion Probe-Window | Standard: quartz; sapphire optional |
| Immersion Probe-Probe Tube Material | Standard: 316 stainless steel; Hastelloy and other materials customizable |
| Software Functions | On-line monitoring, spectrum acquisition, trend display, model research |
| Operating Environment-Operating Temperature | 5 ~ 35 °C |
| Operating Environment-Storage Temperature | -20 ~ 55 °C |