ST50S Transmission Imaging Spectrometer
• Research-grade Raman spectroscopy detection system: 1064nm Confocal Raman microscopy
• Integration of industrial Raman system: Chemical industry Online, Biopharmaceuticals
| Performance Indicators | Parameters | |
| Detector | - | See model table for detailed parameters |
| Optical Parameters | Wavelength Range | 1080nm~1330nm corresponds to 140~1880cm-1 |
| Optical Resolution | 0.35nm, corresponds to 8cm-1(50μm slit) 0.25nm, corresponds to 6cm-1(25μm slit) |
|
| Grating Type | VPH volume holographic transmission grating | |
| Diffraction Efficiency | >85% | |
| Fiber Interface | SMA905 or Ф10mm multi-core optical fiber | |
| Numerical Aperture | 0.25 | |
| Electrical Parameters | Integration Time | 1ms-3600s |
| Data Output Interface | USB or serial port | |
| ADC Bit Depth | 16-bit | |
| Power Supply | DC 5V (±0.5V) | |
| Operating Current | 3A | |
| Operating Temperature | -20°C~60°C | |
| Storage Temperature | -30°C~70°C | |
| Operating Humidity | < 90% RH (non-condensing) | |
| Physical Parameters | Dimensions | 253mm*152mm*93mm |
| Weight | <4kg (including camera) |
| Product Model | ST50S1 | ST50S2 |
| Detector Brand or Model | Hamamatsu Secondary deep cooling InGaAs | AndoriDus InGaAs |
| Number of Pixels | 512*1 | 512*1 |
| Pixel size | 25μm*500μm | 25μm*500μm |
| Cooling Temperature °C | -20 | -80 |
Write your message here and send it to us